water quality monitoring
water quality monitoring
Total phosphorus online water quality analyzer HT-TP-I type

Total phosphorus online water quality analyzer HT-TP-I type

二维码

Products and Applications
Product Details

  Product Overview

  The instrument uses persulfate ammonium molybdate spectrophotometric method to measure the total phosphorus content in water. Peroxysulfate converts all forms of phosphorus into orthophosphate, which reacts with ammonium molybdate to form a blue phosphomolybdic acid. By measuring the absorption of specific wavelengths of light by this blue substance, the total phosphorus content in water can be calculated. The instrument is suitable for online monitoring of various enterprise wastewater discharge outlets, surface water, rivers and lakes, etc.

  Product Features

  ▶ Equipped with automatic fault handling to reduce operation and maintenance workload

  ▶ Independent research and development, supporting personalized customization

  ▶ High precision injection pump, precise and stable quantification

  ▶ Integrated digestion tank, durable, reinforced and sealed

  ▶ High quality pipelines, acid, alkali, and corrosion resistant

  ▶ Experienced operation and maintenance team, worry free after-sales service

  ▶ The measurement principle is the same as the laboratory method, which is convenient for comparison and accurate for testing

  Product parameters

measurement range

0five0mg/L(Can be customized according to customer needs)

indication error

±5

Quantitative lower limit

zero point zero twomg/L

repetitiveness

2%

24-hour low concentration drift

zero point zero onemg/L

24-hour high concentration drift

1%

measurement method

Manual measurement; Interval measurement; Automatic measurement (user selected time point measurement)

Calibration function

Manual calibration; Automatic calibration; Zero point calibration; Calibration of standard solution

Verification function

Manual verification; Automatic verification

signal output

2-channel RS485; 2-channel 4-20mA; 1-channel RS232

communication protocol

HJ212; Modbus

Environmental Requirements

5~40℃

working power supply

AC(220±10%)V,(50±1%)Hz

data storage

At least one year of data, data will not be lost during power outages, and data can be exported from a USB flash drive

power

<170W

power failure protection

Automatically discharge residual liquid after power on and restore to working condition

Instrument size

Length 46cm x Width 36cm x Height 150cm (subject to actual product)

weight

About 60kg

other

Personalized customization (various switch outputs, valve control, water pump control, etc.)



微信扫一扫