water quality monitoring
water quality monitoring
Water quality multi-parameter online monitor, Model HT-WCS-I

Water quality multi-parameter online monitor, Model HT-WCS-I

二维码

Products and Applications
Product Details

  Product Overview

  The water quality multi parameter online monitoring instrument can flexibly combine and select sensors according to user measurement needs, achieving real-time monitoring of multiple water quality parameters on a single machine.

  Optional factors

  PH、 Dissolved oxygen, conductivity, turbidity, suspended solids, temperature ORP、COD、 Ammonia nitrogen, nitrate, fluoride, chloride, chlorophyll, blue-green algae, oil in water, sludge concentration, liquid level, pressure, flow rate, flow rate, etc.

  Product parameters

routineFive parameters

measurement parameters

measurement range

Measuring principle

PH

0~fourteen

Glass electrode method

dissolved oxygen

0~twenty mg/L

fluorescence method

conductivity

0~twenty0000uS/cm

Electrode method

turbidity

0~4000NTU

scattering method

temperature

0~55℃

Platinum thermal resistance method


optionalparameter

measurement parameters

measurement range

Measuring principle

ORP

-2000~2000mV

Platinum electrode method

COD

0~500mg/L
0~500NTU

UV254nm method
546nm  Transmission method turbidity compensation

Ammonia nitrogen, chloride, fluoride

0~ten00mg/L

Ion selective electrode method

chlorophyll

0~500.0 ug/L (chlorophyll-a)
0~ten00ppb  (Jordan Ming WT)

fluorescence method 460nm/680nm

blue-green algae

0~2,000  Kcells/mL

fluorescence method 590nm/650nm

oil in water

0~50/ten0ppm

fluorescence method 360nm/460nm

Sludge concentration

0~5/50g/L

scattering method


  Performance indicators

Conventional five parameters

name

dissolved oxygen

pH

conductivity

turbidity

water temperature

repeatability error

±0.3mg/L

±zero point onepH

±1%

±5%

N/A

zero drift

±0.3  mg/L

N/A

±1%

±3%

N/A

range drift

±0.3  mg/L

N/A

±1%

±5%

N/A

drift

N/A

±zero point onepH

N/A

N/A

N/A

response time

2min

0.5min

0.5min

N/A

N/A

Temperature compensation accuracy

±0.3  mg/L

Within ± zero point onepH

±1%

N/A

N/A

measurement error

N/A

N/A

N/A

N/A

±0.5℃

 

optional parameter

name

ORP

COD

ammonia nitrogen

chlorophyll

blue-green algae

accuracy

±1mV

±7%

±10%

±5%

±5%

resolution

zero point one

zero point zero one

zero point zero one

zero point zero one/0.1

zero point zero one/0.1/1

Response time (s)

15

NC

15

10

10


other

Environmental requirements: Temperature 0~50℃; 10~90%RH

electricity Source: 100~240 V AC, 50 Hz ± 10%

Sensor protection level: IP65 Protection level

Communication: RS485 Modbus/HJ212 (supports customization)

Abnormal alarms and power outages will not result in data loss

Touch screen display and command input

Automatic restoration of working status after abnormal reset and power outage when power is restored

Customizable integration of various functions